H₂S IP 570 Proficiency Test Scheme
Part number SA4032-0
Produced in general accordance with the principles of ISO 17034 and ISO 33405, the Multi Test Certified Reference Material (CRM) Gas Oil provides certified values for multiple test methods and has been characterised using a network of competent laboratories to provide international traceability. It is a highly cost effective solution to help meet verification requirements for audited laboratories and those which comply with ISO 17025.
The Multi Test CRM Gas Oil is supplied with full documentation including Safety Data Sheet (SDS), in tamper evident security packaging. It comes with a 1.5 year shelf life from date of manufacture.
Pack size: 500 mL.
Part Number: | SP9000-0 |
---|---|
Methods: | ASTM D2162, ASTM D4052, ASTM D86, ASTM D93, IP 219, IP 309 |
The Gas Oil Multi Test Certified Reference Material allows routine monitoring of instrument performance and is also particularly helpful in training personnel on new equipment and test methods. The purchase of Multi Test CRMs is recommended with new instrumentation to assist with installation and commissioning.
Incorrect determination of a result can have far reaching financial and safety implications. The growing use of automated instrumentation increases the possibility that incorrect results may go undetected and in addition the early indication of bias is a useful warning to minimise production ‘giveaway’.
Seta Certified Reference Materials are unique because, unlike most other Certified Reference Material that are only suitable for validating a specific parameter/value, Seta CRMs enable a user to validate different tests and instrumentation using the same sample material.
Certification: the Multi Test Certified Reference Material (CRM) Gas Oil is supplied with a certificate of measurement which gives the certified value for each test method covered by the material.
Test Name | Test Methods | Certified Value |
---|---|---|
Kinematic Viscosity @ 40 °C | ASTM D2162 | 2.923 mm2/s (cSt) |
Dynamic Viscosity @ 40 °C | ASTM D2162 | 2.386 mPa s (cP) |
Density @ 15 °C, Digital Density Meter | ASTM D4052 | 0.83373 g/mL |
Cloud Point | IP 219 | -3.1 °C |
CFPP | IP 309 | -10.3 °C |
Distillation IBP | ASTM D86 | 170.2 °C |
5% v/v recovered | ASTM D86 | 196.7 °C |
10% v/v recovered | ASTM D86 | 209.6 °C |
20% v/v recovered | ASTM D86 | 231.2 °C |
30% v/v recovered | ASTM D86 | 248.4 °C |
40% v/v recovered | ASTM D86 | 262.6 °C |
50% v/v recovered | ASTM D86 | 275.2 °C |
60% v/v recovered | ASTM D86 | 287.3 °C |
70% v/v recovered | ASTM D86 | 299.9 °C |
80% v/v recovered | ASTM D86 | 314.6 °C |
90% v/v recovered | ASTM D86 | 334.7 °C |
95% v/v recovered | ASTM D86 | 351.0 °C |
FBP | ASTM D86 | 360.5 °C |
Flash Point, PMCC | ASTM D93 Procedure A | 68.2 °C |
Method | Details |
---|---|
ASTM D2162 | Standard Practice for Basic Calibration of Master Viscometers and Viscosity Oil Standards |
ASTM D4052 | Standard Test Method for Density, Relative Density, and API Gravity of Liquids by Digital Density Meter |
ASTM D86 | Standard Test Method for Distillation of Petroleum Products at Atmospheric Pressure |
ASTM D93 | Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester |
IP 219 | Determination of cloud point |
IP 309 | Diesel and domestic heating fuels - Determination of cold filter plugging point |
Stanhope-Seta provide direct support by phone and email.